Measurement uncertainty expresses the doubt attached to every measurement result. It gives the interval within which, with a known probability, the true value of what is being measured lies.

For example, the probe in cold room CF-03 reads 4.0 °C and its calibration certificate states an uncertainty of ±0.3 °C at a confidence level of approximately 95%. This means that, with that probability, the true temperature is between 3.7 °C and 4.3 °C.

Uncertainty matters most near the limits. If the cold room must stay below 5 °C and the probe reads 4.8 °C, an uncertainty of ±0.3 °C means compliance cannot be confirmed. That is why it is taken into account when deciding whether an instrument is fit for its intended use.